Title of Article

Capabilities of detecting low-concentration high-toxic bromate ions by chromatograph «844 UV/VIS COMPACT IC»


Issue
1
Date
2009

Section
CHEMISTRY

Article type
scientific article
UDC
543.544.5 : 546.145
Pages
53-57
Keywords
bromate ions, low concentrations, detection, ion chromatography


Authors
Shlyapunova E.V.
Nizhegorodskiy gosuniversitet im. N.I. Lobachevskogo

Sergeev G.M.
Nizhegorodskiy gosuniversitet im. N.I. Lobachevskogo


Abstract
We have studied the capabilities of the chromatograph «844 UV/VIS Compact IC» to detect low-concentration high-toxic bromate ions that are by-products of drinking water ozonation. The optimal conditions for the analysis have been found. The results obtained (the ion detection limit is 0.5 ?g/dm3; the total error is 5–10% in the range of concentrations (1–20) ?g/dm3) correspond to existing international standards.